Subj: Measuring 1 of 3 Rs
Path: HB9ON<DK0WUE<PD0LPM<VE3CGR<VE3TOK<IR0AAB<I0OJJ<GB7CIP
Sent: 250220/1012Z @:GB7CIP.#32.GBR.EURO #:26450 [Caterham Surrey GBR]

>From g8mny%gb7cip.#32.gbr.euro@i0ojj.ampr.org Thu Feb 20 11:13:33 2025
Received: from i0ojj.ampr.org by i0ojj.ampr.org (JNOS2.0p.6) with SMTP
	id AA17022733 ; Thu, 20 Feb 2025 11:13:33 +0100
Message-Id: <26450_GB7CIP@gb7cip.bbs>
>From: g8mny@gb7cip.#32.gbr.euro
X-JNOS-User-Port: Telnet   (gb7cip @ 77.104.162.230)  -> Sending message

From: G8MNY@GB7CIP.#32.GBR.EURO
To  : TECH@WW

By G8MNY                                           (new Mar 06)
(8 Bit ASCII graphics use code page 437 or 850, Terminal Font)
I recently bought an old LAB DMM (Datron 1061a) meter. It has an unusual
feature, the ability to measure one Resistor R1 in a triangle of 3 (R2 & R3)
while all connected.

ÚÄÄÄÄÄÄÄÄ¿
³    HighÃÄÄÄÄÂÄÄÄ¿
³        ³    ³  R2
³     Rx>³   R1   ÃÄÄ¿
³ DMM    ³    ³  R3  ³
³     LowÃÄÄÄÄÁÄÄÄÙ  ³
³        ³           ³
³   GuardÃÄÄÄÄÄÄÄÄÄÄÄÙ
ÀÄÄÄÄÄÄÄÄÙ

As the DMM has uProcessor you might expect the Guard line to be connected to
the Low then High & then disconnected, to give 3 values of Rx, then with 3
equations & 3 unknowns all the Rs can be calculated....

1/ Guard                  R1 + R2
   connected         Rx = ÄÄÄÄÄÄÄ
   to Low                 R1 x R2

2/ Guard                  R1 + R3
   connected         Rx = ÄÄÄÄÄÄÄ
   to High                R1 x R3

3/ Guard                  R1 + R2 + R3
   not               Rx = ÄÄÄÄÄÄÄÄÄÄÄÄ
   connected              R1 x (R2+R3)

But that is not the purpose of the Guard line in a sensitive measuring circuit.

To measure the resistance the DMM actually uses an accurate constant current
source across Rx & measures the voltage to get the R1 value. But with R2 + R3
in circuit this will give a wrong reading as some of the accurate current will
flow through R2 + R3. But on this meter the Guard line has another current
generator with its control arranged to make the Guard the same potential as the
Low...

ÚÄÄÄÄÄÄÄÄÄÄÄÄ¿
³        HighÃÄÄÄÄÂÄÄÄ¿
³          / ³    ³  R2
³        DVM ³   R1   ÃÄÄ¿
³          \ ³    ³  R3  ³
³  Ref I> LowÃÄÄÄÄÁÄÄÄÙ  ³
³       /³_ /³   No /³\  ³
³  ÚÄÄ<' ³_  ³  Voltage  ³
³ \³/  `\³  \³      \³/  ³
³ Stray I> GuÃÄÄÄÄÄÄÄÄÄÄÄÙ
ÀÄÄÄÄÄÄÄÄÄÄÄÄÙ

So as long as the Guard stray current generator can put the error current
through R2 the Reading of Rx = R1 & be correct.


See my tech bul "Lab Measuring DMM".

Why Don't U send an interesting bul?

73 De John, G8MNY @ GB7CIP